Evaluation of the gamma dose distribution comparison method
Washington University in St. Louis · University of Florida
Abstract
The gamma tool was developed to quantitatively compare dose distributions, either measured or calculated. Before computing gamma, the dose and distance scales of the two distributions, referred to as evaluated and reference, are renormalized by dose and distance criteria, respectively. The renormalization allows the dose distribution comparison to be conducted simultaneously along dose and distance axes. The gamma quantity, calculated independently for each reference point, is the minimum distance in the renormalized multidimensional space between the evaluated distribution and the reference point. The gamma quantity degenerates to the dose-difference and distance-to-agreement tests in shallow and very steep…