articleCirculationJan 27, 2004BRONZE OA

Localized Hypersensitivity and Late Coronary Thrombosis Secondary to a Sirolimus-Eluting Stent

Centro Studi GISED

PubMed
Indexed incrossrefpubmed

Abstract

Background

The US Food and Drug Administration recently issued a warning of subacute thrombosis and hypersensitivity reactions to sirolimus-eluting stents (Cypher). The cause and incidence of these events have not been determined. METHODS AND RESULTS: We present findings of a 58-year-old man who died of late stent thrombosis 18 months after receiving 2 Cypher stents for unstable angina. Although angiographic and intravascular ultrasound results at 8 months demonstrated the absence of neointimal formation, vessel enlargement was present. An autopsy showed aneurysmal dilation of the stented arterial segments with a severe localized hypersensitivity reaction consisting predominantly of T lymphocytes and eosinophils.

Conclusions

The known pharmacokinetic elution profile of Cypher stents and the presence of polymer fragments surrounded by giant cells and eosinophils suggest that a reaction to the polymer may have caused late stent thrombosis. Careful long-term follow-up of patients with vessel enlargement after Cypher stent placement is recommended.

Citation impact

1,465
total citations
FWCI
99.07
Percentile
100%
References
29
Citations per year

Authors

10

Topics & keywords

Keywords
  • Medicine
  • Sirolimus
  • Stent
  • Thrombosis
  • Intravascular ultrasound
  • Autopsy
  • Hypersensitivity reaction
  • Drug-eluting stent
UN Sustainable Development Goals
  • Good health and well-being
No related works found for this paper.