Visual Acuities “Hand Motion” and “Counting Fingers” Can Be Quantified with the Freiburg Visual Acuity Test
Abstract
Examined were 100 eyes of 100 patients with various eye diseases (e.g., diabetic retinopathy, ARMD), covering a range of VAs from LP to decimal 0.32. The FrACT optotypes were presented on a 17-inch LCD monitor with random orientation. After extensive training, two ETDRS and FrACT measurements were obtained. The testing distance was 50 or 100 cm.
ETDRS and FrACT coincided closely for VA > or = 0.02 (n = 80). ETDRS measures were successfully obtainable down to CF (at 30 cm; test-retest averaged over all patients, coefficient of variation [CV](ETDRS) = 9% +/- 8%), and FrACT provided reproducible measurements down to HM (test-retest CV(FrACT) =12% +/- 11%). For CF (n = 6), both ETDRS and FrACT resulted in a mean VA of 0.014 +/- 0.003 (range, 0.01-0.02). The VA results of FrACT for HM (n = 12) were 0.005 +/- 0.002 (range, 0.003-0.009); the individual values were highly reproducible. No results were obtainable for LP (n = 2).