articleJournal of Cataract & Refractive SurgeryFeb 26, 2008Closed access

Sources of error in intraocular lens power calculation

University of Groningen · Dialyse Centrum Groningen

PubMed
Indexed incrossrefpubmed

Abstract

Setting

AMO Groningen BV, Groningen, The Netherlands.

Methods

Means and standard deviations (SDs) of parameters that influence refractive outcomes were taken or derived from the published literature to the extent available. To evaluate their influence on refraction, thick-lens ray tracing that allowed for asphericity was used. The numerical partial derivative of each parameter with respect to spectacle refraction was calculated. The product of the partial derivative and the SD for a parameter equates to its SD, expressed as spectacle diopters, which squared is the variance. The error contribution of a parameter is its variance relative to the sum of the variances of all parameters.

Citation impact

649
total citations
FWCI
10.49
Percentile
100%
References
27
Citations per year

Authors

1

Topics & keywords

Keywords
  • Dioptre
  • Refractive error
  • Refraction
  • Mathematics
  • Intraocular lens
  • Optics
  • Lens (geology)
  • Spherical aberration
No related works found for this paper.