Analytical Validation of a High-Sensitivity Cardiac Troponin T Assay
University Hospital Heidelberg · Roche Pharma AG (Germany) · +2 more institutions
Indexed incrossrefpubmed
Abstract
Background
We report the development of a novel high-sensitivity cardiac troponin T (hs-cTnT) assay, a modification of the Roche fourth-generation cTnT assay, and validation of the analytical performance of this assay.
Methods
Validation included testing of analytical sensitivity, specificity, interferences, and precision. We established the 99th percentile cutoff from healthy reference populations (n = 616). In addition, we studied differences in time to a positive result when using serial measurements of hs-cTnT vs cTnT in patients with a confirmed diagnosis of non-ST elevation myocardial infarction (non-STEMI).
Citation impact
1,075
total citations
- FWCI
- 34.49
- Percentile
- 100%
- References
- 27
Citations per year