articleMedical PhysicsAug 17, 2009BRONZE OA

Task Group 142 report: Quality assurance of medical acceleratorsa)

Washington University in St. Louis · Hackensack University Medical Center · +10 more institutions

PubMed
Indexed incrossrefpubmed

Abstract

The task group (TG) for quality assurance of medical accelerators was constituted by the American Association of Physicists in Medicine's Science Council under the direction of the Radiation Therapy Committee and the Quality Assurance and Outcome Improvement Subcommittee. The task group (TG-142) had two main charges. First to update, as needed, recommendations of Table II of the AAPM TG-40 report on quality assurance and second, to add recommendations for asymmetric jaws, multileaf collimation (MLC), and dynamic/virtual wedges. The TG accomplished the update to TG-40, specifying new test and tolerances, and has added recommendations for not only the new ancillary delivery technologies but also for imaging…

Citation impact

1,584
total citations
FWCI
58.04
Percentile
100%
References
83
Citations per year

Authors

13

Topics & keywords

Keywords
  • Quality assurance
  • Task group
  • Medical physics
  • Medical physicist
  • Medical imaging
  • Linear particle accelerator
  • Tomotherapy
  • Computer science