Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner
TUM Klinikum · Technical University of Munich · +4 more institutions
Abstract
The NEMA NU 2-2007 protocol was followed for studying the PET performance. The following measurements were performed: spatial resolution; scatter fraction, count losses, and randoms; sensitivity; accuracy of the correction for count losses and randoms; and image quality. The quality control manual of the American College of Radiology was followed for studying the MR performance. The following measurements were performed: geometric accuracy, spatial resolution, low-contrast detectability, signal-to-noise ratio, static field (B(0)) homogeneity, radiofrequency field (B(1)) homogeneity, and radiofrequency noise.
An average spatial resolution of 4.3 mm in full width at half maximum was measured at 1 cm offset from the center of the field of view. The system sensitivity was 15.0 kcps/MBq along the center of the scanner. The scatter fraction was 37.9%, and the peak noise-equivalent count rate was 184 kcps at 23.1 kBq/mL. The maximum absolute value of the relative count rate error due to dead-time losses and randoms was 5.5%. The average residual error in scatter and attenuation correction was 12.1%. All MR parameters were within the tolerances defined by the American College of Radiology. B(0) inhomogeneities below 1 ppm were measured in a 120-mm radius. B(1) homogeneity and signal-to-noise ratio were equivalent to those of a standard MR scanner. No radiofrequency interference was detected.
Citation impact
- FWCI
- 51.00
- Percentile
- 100%
- References
- 26
Authors
8Topics & keywords
- Scanner
- Nuclear medicine
- Image quality
- Image resolution
- Repeatability
- Physics
- Optics
- Mathematics