Reduction of eddy‐current‐induced distortion in diffusion MRI using a twice‐refocused spin echo
Massachusetts General Hospital · Siemens (Germany)
Abstract
Image distortion due to field gradient eddy currents can create image artifacts in diffusion-weighted MR images. These images, acquired by measuring the attenuation of NMR signal due to directionally dependent diffusion, have recently been shown to be useful in the diagnosis and assessment of acute stroke and in mapping of tissue structure. This work presents an improvement on the spin-echo (SE) diffusion sequence that displays less distortion and consequently improves image quality. Adding a second refocusing pulse provides better image quality with less distortion at no cost in scanning efficiency or effectiveness, and allows more flexible diffusion gradient timing. By adjusting the timing of the diffusion…