Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT
University of North Carolina at Chapel Hill · University of Michigan–Ann Arbor
Abstract
Cone beam computed tomography (CBCT), which provides a lower dose, lower cost alternative to conventional CT, is being used with increasing frequency in the practice of oral and maxillofacial radiology. This study provides comparative measurements of effective dose for three commercially available, large (12'') field-of-view (FOV), CBCT units: CB Mercuray, NewTom 3G and i-CAT.
Thermoluminescent dosemeters (TLDs) were placed at 24 sites throughout the layers of the head and neck of a tissue-equivalent human skull RANDO phantom. Depending on availability, the 12'' FOV and smaller FOV scanning modes were used with similar phantom positioning geometry for each CBCT unit. Radiation weighted doses to individual organs were summed using 1990 (E(1990)) and proposed 2005 (E(2005 draft)) ICRP tissue weighting factors to calculate two measures of whole-body effective dose. Dose as a multiple of a representative panoramic radiography dose was also calculated.
Citation impact
- FWCI
- 42.58
- Percentile
- 100%
- References
- 20
Authors
4Topics & keywords
- Imaging phantom
- Nuclear medicine
- Dosimetry
- Medicine
- Cone beam computed tomography
- Radiography
- Field size
- Effective dose (radiation)