Evaluation of the Infinium Methylation 450K technology

Université Libre de Bruxelles

PubMed
Indexed incrossrefpubmed

Abstract

Aims

Studies of DNA methylomes hold enormous promise for biomedicine but are hampered by the technological challenges of analyzing many samples cost-effectively. Recently, a major extension of the previous Infinium HumanMethylation27 BeadChip® (Illumina, Inc. CA, USA), called Infinium HumanMethylation450 (Infinium Methylation 450K; Illumina, Inc. CA, USA) was developed. This upgraded technology is a hybrid of two different chemical assays, the Infinium I and Infinium II assays, allowing (for 12 samples in parallel) assessment of the methylation status of more than 480,000 cytosines distributed over the whole genome. In this article, we evaluate Infinium Methylation 450K on cell lines and tissue samples, highlighting some of its advantages but also some of its limitations. In particular, we compare the methylation values of the Infinium I and Infinium II assays. MATERIALS & METHODS: We used Infinium Methylation 450K to profile: first, the well-characterized HCT116 wild-type and double-knockout cell lines and then, 16 breast tissue samples (including eight normal and eight primary tumor samples). Absolute methylation values (β-values) were extracted with the GenomeStudio™ software and then subjected to detailed analysis.

Results

While this technology appeared highly robust as previously shown, we noticed a divergence between the β-values retrieved from the type I and type II Infinium assays. Specifically, the β-values obtained from Infinium II probes were less accurate and reproducible than those obtained from Infinium I probes. This suggests that data from the type I and type II assays should be considered separately in any downstream bioinformatic analysis. To be able to deal with the Infinium I and Infinium II data together, we developed and tested a new correction technique, which we called 'peak-based correction'. The idea was to rescale the Infinium II data on the basis of the Infinium I data. While this technique should be viewed as an approximation method, it significantly improves the quality of Infinium II data.

No related works found for this paper.