Pacemakers and Implantable Cardiac Defibrillators: Software Radio Attacks and Zero-Power Defenses
University of Washington · University of Massachusetts Amherst · +1 more institution
Abstract
Our study analyzes the security and privacy properties of an implantable cardioverter defibrillator (ICD). Introduced to the U.S. market in 2003, this model of ICD includes pacemaker technology and is designed to communicate wirelessly with a nearby external programmer in the 175 kHz frequency range. After partially reverse-engineering the ICD's communications protocol with an oscilloscope and a software radio, we implemented several software radio-based attacks that could compromise patient safety and patient privacy. Motivated by our desire to improve patient safety, and mindful of conventional trade-offs between security and power consumption for resource-constrained devices, we introduce three new…
Citation impact
- FWCI
- 23.48
- Percentile
- 100%
- References
- 34
Authors
9Topics & keywords
- Computer security
- Programmer
- Computer science
- Software
- First responder
- Cryptography
- Protocol (science)
- Radio frequency