articleIEEE Transactions on Nuclear ScienceOct 1, 2002Closed access

Medipix2: A 64-k pixel readout chip with 55-/spl mu/m square elements working in single photon counting mode

European Organization for Nuclear Research · National Institute for Subatomic Physics · +2 more institutions

Indexed incrossref

Abstract

The Medipix2 chip is a pixel-detector readout chip consisting of 256 /spl times/ 256 identical elements, each working in single photon counting mode for positive or negative input charge signals. Each pixel cell contains around 500 transistors and occupies a total surface area of 55 /spl mu/m /spl times/ 55 /spl mu/m. A 20-/spl mu/m wide octagonal opening connects the detector and the preamplifier input via bump bonding. The preamplifier feedback provides compensation for detector leakage current on a pixel by pixel basis. Two identical pulse height discriminators are used to create a pulse if the preamplifier output falls within a defined energy window. These digital pulses are then counted with a 13-b…