articleIEEE Transactions on Biomedical EngineeringJun 14, 2016Closed access

Cuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring

Sharif University of Technology

PubMed
Indexed incrossrefpubmed

Abstract

Methods

The proposed framework estimates the BP values through processing vital signals and extracting two types of features, which are based on either physiological parameters or whole-based representation of vital signals. Finally, the regression algorithms are employed for the BP estimation. Although the proposed algorithm works reliably without any need for calibration, an optional calibration procedure is also suggested, which can improve the system's accuracy even further.

Results

The proposed method is evaluated on about a thousand subjects using the Association for the Advancement of Medical Instrumentation (AAMI) and the British Hypertension Society (BHS) standards. The method complies with the AAMI standard in the estimation of DBP and MAP values. Regarding the BHS protocol, the results achieve grade A for the estimation of DBP and grade B for the estimation of MAP.

Citation impact

596
total citations
FWCI
15.75
Percentile
100%
References
37
Citations per year

Authors

4

Topics & keywords

Keywords
  • Blood pressure
  • Calibration
  • Medical instrumentation
  • Computer science
  • Estimation
  • Algorithm
  • Artificial intelligence
  • Medicine
No related works found for this paper.