articleJournal of Proteome ResearchMay 14, 2025HYBRID OA

Technical Evaluation of Plasma Proteomics Technologies

Morgridge Institute for Research · Utrecht University

PubMed
Indexed incrossrefpubmed

Abstract

Plasma proteomics technologies are rapidly evolving and of critical importance to the field of biomedical research. Here, we report a technical evaluation of six notable plasma proteomics technologies─unenriched (Neat), acid depletion, PreOmics ENRICHplus, Mag-Net, Seer Proteograph XT, and Olink Explore HT. The methods were compared on proteomic depth, reproducibility, linearity, tolerance to lipid interference, and limit of detection/quantification. In total, we performed 618 LC-MS/MS experiments and 93 Olink Explore HT assays. The Seer method achieved the greatest proteomic depth (∼4500 proteins detected), while Olink detected ∼2600 proteins. Other MS-based methods ranged from ∼500-2200 proteins detected. In…

Citation impact

44
total citations
FWCI
24.81
Percentile
100%
References
45
Citations per year

Authors

6

Topics & keywords

Keywords
  • Proteomics
  • Computational biology
  • Computer science
  • Chemistry
  • Data science
  • Biology
  • Biochemistry
No related works found for this paper.

Funding