articleAnalytical ChemistryJan 2, 2026Closed access

Stepped-Type Photoacoustic Resonator for Simultaneous Dual-Gas Detection

Dalian University of Technology · Electric Power Research Institute · +1 more institution

PubMed
Indexed incrossrefpubmed

Abstract

A stepped-type resonator-based photoacoustic (PA) sensor is proposed for simultaneous dual-gas detection. The sensor employs a dual-resonance architecture, with a stepped resonant tube capable of acoustic amplification at two different frequencies. An optimized design of the PA resonator response is conducted using finite element analysis. When the amplitudes of the two resonance peaks are high and the relative signal strength is also high, the simulated values of the two resonance frequencies are 2370 and 4220 Hz, respectively. A dual-channel synchronous digital lock-in amplifier is employed to achieve simultaneous detection in the two channels. Two lasers for methane and acetylene detection are modulated…

Citation impact

6
total citations
FWCI
44.93
Percentile
100%
References
60
Too recent for citation history.

Authors

7

Topics & keywords

Keywords
  • Resonator
  • Photoacoustic imaging in biomedicine
  • Amplifier
  • Laser
  • Resonance (particle physics)
  • Photoacoustic spectroscopy
  • Time delay and integration
  • Detection limit
UN Sustainable Development Goals
  • Affordable and clean energy
No related works found for this paper.

Funding